top of page

FPI AQMS-600 (MTI.AQMS.FPI.600) nitrogen oxides (NO-NO2–NOx) analyzer utilizes chemiluminescence technology indicated by US EPA as federal reference meth- od for monitoring on multiple forms of nitrogen oxides.

NOx (AQMS-600)

SKU: MTI.AQMS.FPI.600
  • • Principle: Chemiluminiscence

    • Standard Range: (0-500)ppb, expandable

    • Zero Noise: ≤0.2ppb(RMS)

    • Span Noise: ≤2.5ppb(RMS)

    • Display: Digital

    • Lower Detectable Limit: ≤0.4ppb(RMS)

    • Zero Drift: ±1ppb/24h

    • Span Drift: Span drift(20%F.S): ±5ppb Span drift(80%F.S): ±10ppb

    • Linearity: <1%F.S.

    • Precision: Precision(20%F.S): ≤2ppb Precision(80%F.S): ≤4ppb

    • Response Time: T90<60s

    • Sample Flow Rate: ( 500±50)sccm

    • Data Transmission: 2 channel analog(4~20) mA; 12 digital input/output; 4-way relay output;

    • Calibration: Multi-point calibrator

    • Output: RS232/RS485/Ethernet

    • Operating Temperature: 0~40°C

    • Operating Humidity: 0~95%RH(No condensation)

    • Power Requirement: (220±10%)V AC/(45~55)Hz, power≤350W

    • Dimensions and Weight: 178(H)x432(W)x609(D)mm, 22kg

WhatsApp
bottom of page