The Turbidity Analyzer is connected directly via RS485 communication interface, providing simple, reliable, cost-saving process data with remote monitoring, calibration, configuration and diagnostic capabilities. Housing in a robust IP68 proof enclosure, 1500 N tensile strength Kevlar reinforced cable, up to 1.2 km digital data transmission, the analyzer is ideally used in water/wastewater industry.
Turbidity Sensor (SMR10-2)
Output Signal
• Output Signal: RS485 (Modbus RTU protocol), 19,200 bps, 8 data bits, no parity, 1 stop bit; 4~20 mA
• Data Resolution: 16 bits (0.001% FS)
• Surge Protection: 1,500 VDC
• Power: 5~12 VDC, 92 mA
• Protection: Polarity, Overload, Short circuit
• This measuring device is a Nephelometric Turbidity Unit (NTU) according to ISO7027 (Formazin calibration solution)
• Safety: CE, FCC
Turbidity
• Measurement range: SMR10-2: 0~10/100/1,000/4,000 NTU (depending on sample), auto range SMR10-4: 0~10/100/1,000/4,000/10,000 NTU (depending on sample), auto range
• Accuracy ±2% measured value (0~1,000 NTU) ; ±5% measured value (1,000~10,000 NTU)
• Resolution: 0.001 NTU
• Repeatability: ±1% measured value
• Light source: LED 880 nm
• Process flow rate: 0.1~10 m/s
• Operation Pressure: Max.10 Kgf/cm²
• Operating Temperature: 0~60 °C
• Response Time: 3 secs
• Calibration Interval: Typical 24 months
• Protection: IP68
• Connection: M16X1.5 plug fixed cable, M12 connector, 5 pin • Housing Material: SS316L ; Titanium
• Cable: Kevlar reinforced PUR cable, 1500N tensile strength
• Dimension: ø 22Χ130 mm (no wiper) ; ø 39Χ186.2mm (with wiper)
• Weight: analyzer: approx. 500g (SS316L); 400 g (Titanium) ;
• Cable: 80 g/m