Nitrate ion selective electrode is an effective method to measure the content of nitrate ion in the sample. Nitrate ion selective electrodes are also often used in online instruments, such as the monitoring of industrial online nitrate ion content. Nitrate ion selective electrode has the advantages of simple measurement, fast and accurate response, and can be used with pH meter, ion meter and online nitrate ion analyzer.
Nitrat (W610-A)
• Slope: 55 ± 3mv
• Repeatability: ± 2%
• Interference: ClO4 -1, I-1, CN-1, BF4 -1,Cl– , NO2– , organic acid ion \
• Temperature range: 0-50°C
• Pressure range: 0-2 atm
• Responding speed 95% response within 30 seconds
• Storage: Long term: dry storage; Short-term: stored in diluted nitrate standard solution
• Measuring range: 0.5ppm –62,000ppm
• pH range: 2.5-11PH
• Output impedance: 20-50MΩ
• Maintenance and cleaning: Deionized water and treated with diluted nitrate standard solution
• Temperature compensation: 10k/22k//2.252K/PT1000/PT100 optional
• Work online: Can
• Equipotential point: 10 ppm NO3-
• Reference electrode: Double salt bridge, single hole ceramic liquid junction
• Material: PMMA+PVC
• Connector: terminal